Cu+/Cu2+

Applications
- Industrial wastewater
- Galvanic processes
- Process water
- Industrial sewage treatment plants
- Ultrapure water treatment
Description
The determination ranges of the copper analyzer vary from trace μg/L to 150 mg/L using the (optional) internal dilution module.
Method
Copper (I) ions form an orange-colored complex with the bathocuproine in a buffered solution at pH 4,5. Any copper (II) ions present in the water sample are previously reduced to copper (I) ions before the complex is formed. The absorbance intensity is proportional to the copper concentration in the sample and is determined at 470 nm.
Reagents
Standard solutions
Features
Range (Ø 26 mm) |
1000 ppb |
Range (Ø 16 mm) |
3000 ppb |
Range (with dilution) |
150 ppm |
Wavelength | 470 nm |
Number of reagents | 3 |
Min. analysis time | 8 minutes |
Technical Specifications
Model | 3S-CL |
Sampling Mode | Batch, with freely settable frequency. |
Sample | Pressure: Atmospheric Temperature: 5 ‐ 45°C (41 ‐ 113°F) Flow Rate: 80 to 500 mL/min Connection: 6 mm (¼-in.) |
Drain |
Pressure-free drain Connection: 12 mm (½-in.) |
Input Streams | 2 |
Material | Epoxy-coated stainless steel AISI 304 |
Dimensions (H x W x D) | 604 x 380 x 210 mm (23.6 x 14.8 x 8.2 in) |
Weight | 20 kg (44 lbs) |
Power Supply |
Input Voltage: 115 VAC,230 VAC,24 VDC Power consumption: max. 80 W |
Output signals |
2 x 4-20 mA analog outputs Modbus via RS485,Ethernet |
Alarms | 2 SPDT programmable voltage-free relays |
Digital Input | None, Online, Start Extra, Skip Idle, Emergency Stop |
Ambient Temperature | 5 - 45 °C (41 - 113 °F) |
Ambient Humidity | 10 - 90 % RH (non-condensing) |
Protection Grade | IP54 (indoor only, outdoor use possible with external cabinet, not included) |