Copper

Cu+/Cu2+

Applications

  • Industrial wastewater
  • Galvanic processes
  • Process water
  • Industrial sewage treatment plants
  • Ultrapure water treatment
Copper

Description

The determination ranges of the copper analyzer vary from trace μg/L to 150 mg/L using the (optional) internal dilution module.

Method

Copper (I) ions form an orange-colored complex with the bathocuproine in a buffered solution at pH 4,5. Any copper (II) ions present in the water sample are previously reduced to copper (I) ions before the complex is formed. The absorbance intensity is proportional to the copper concentration in the sample and is determined at 470 nm.

Features

Measured parameter Cu+/Cu2+

Range

(Ø 26 mm) 50 - 1000 ppb
(Ø 16 mm) 100 - 3000 ppb
(with dilution) 150 ppm

Reproducibility*

* if two values take whichever is greater.
(Ø 26 mm) ± 20 ppb / ± 5%
(Ø 16 mm) ± 50 ppb / ± 5%
Limit of detection 50 ppb
Wavelength 470 nm
Demineralized water Not required
Number of reagents 3
Min. analysis time 9 minutes
Default analysis time 15 minutes

Reagent consumption**

** per month, at the default analysis frequency.
(Ø 26 mm) 2 L
(Ø 16 mm) 1.5 L

Technical Specifications

Model 3S-CL
Sampling Mode Batch, with freely settable frequency.
Sample Pressure: Atmospheric
Temperature: 5 ‐ 45°C (41 ‐ 113°F)
Flow Rate: 80 to 500 mL/min
Connection: 6 mm (¼-in.)
Drain Pressure-free drain
Connection: 12 mm (½-in.)
Input Streams 2
Material Epoxy-coated stainless steel AISI 304
Dimensions (H x W x D) 604 x 380 x 210 mm (23.6 x 14.8 x 8.2 in)
Weight 20 kg (44 lbs)
Power Supply Input Voltage: 115 VAC,230 VAC,24 VDC
Power consumption: max. 80 W
Output signals 2 x 4-20 mA analog outputs
Modbus via RS485,Ethernet
Alarms 2 SPDT programmable voltage-free relays
Digital Input None, Online, Start Extra, Skip Idle, Emergency Stop
Ambient Temperature 5 - 45 °C (41 - 113 °F)
Ambient Humidity 10 - 90 % RH (non-condensing)
Protection Grade IP54 (indoor only, outdoor use possible with external cabinet, not included)