Turbidity

Applications

  • Drinking water
  • Industrial wastewater
  • Process water
  • Pollution monitoring of surface waters
Turbidity

Description

The probe is an online sensor for the determination of water turbidity. The probe is available in various ranges, up to 1000 NTU.

Method

The probe measure turbidity in water samples via light scattering at 90°.

Reagents

No reagents required.

Features

Range 0.1 - 400 NTU
Reproducibility <5% or 0.3 NTU

Technical Specifications

Model 3S-TRB
Sampling Mode Continuous.
Sample Pressure: Atmospheric
Temperature: 2 ‐ 50°C (36 ‐ 122 °F)
Flow Rate: 80 to 500 mL/min
Connection: 6 mm (¼-in.)
Drain Pressure-free drain
Connection: 12 mm (½-in.)
Input Streams 1
Cabinet Material Stainless steel 316L.
Dimensions (H x W x D) 156 x 30 x 30 mm (6.1 x 1.2 x 1.2 in)
Weight 1 kg (2 lbs)
Power Supply Input Voltage: 5 VDC,12 VDC
Power consumption: max. 1 W
Output signals Modbus via RS485
Alarms No alarm relays.
Digital Input None
Ambient Temperature 5 - 50 °C (41 - 122 °F)
Ambient Humidity 100 % RH
Protection Grade IP68 (indoor only, outdoor use possible with external cabinet, not included)

Documentation

Related Accessories