Zinc

Zn2+

Applications

  • Drinking water
  • Industrial waste water
  • Galvanic industry
  • Effluent water
  • Surface water
  • Boilers and cooling towers
Zinc

Description

The determination ranges of the zinc analyzer vary from trace μg/L to 125 mg/L Zn2+ using the (optional) internal dilution module.

Method

Zinc reacts with the reagent zincon in a buffered alkaline solution to form a blue complex. The absorbance intensity is proportional to the zinc concentration in the sample and is measured at 660 nm.

Features

Measured parameter Zn2+

Range

(Ø 26 mm) 0.01 - 1 ppm
(Ø 16 mm) 0.02 - 2.5 ppm
(with dilution) 125 ppm

Reproducibility*

* if two values take whichever is greater.
(Ø 26 mm) ± 10 ppb / ± 5%
(Ø 16 mm) ± 20 ppb / ± 5%
Limit of detection 0.01 ppm
Wavelength 660 nm
Demineralized water Not required
Number of reagents 3
Min. analysis time 5 minutes
Default analysis time 15 minutes

Reagent consumption**

** per month, at the default analysis frequency.
(Ø 26 mm) 1.6 L
(Ø 16 mm) 1.2 L

Technical Specifications

Model 3S-CL
Sampling Mode Batch, with freely settable frequency.
Sample Pressure: Atmospheric
Temperature: 5 ‐ 45°C (41 ‐ 113°F)
Flow Rate: 80 to 500 mL/min
Connection: 6 mm (¼-in.)
Drain Pressure-free drain
Connection: 12 mm (½-in.)
Input Streams 2
Material Epoxy-coated stainless steel AISI 304
Dimensions (H x W x D) 604 x 380 x 210 mm (23.6 x 14.8 x 8.2 in)
Weight 20 kg (44 lbs)
Power Supply Input Voltage: 115 VAC,230 VAC,24 VDC
Power consumption: max. 80 W
Output signals 2 x 4-20 mA analog outputs
Modbus via RS485,Ethernet
Alarms 2 SPDT programmable voltage-free relays
Digital Input None, Online, Start Extra, Skip Idle, Emergency Stop
Ambient Temperature 5 - 45 °C (41 - 113 °F)
Ambient Humidity 10 - 90 % RH (non-condensing)
Protection Grade IP54 (indoor only, outdoor use possible with external cabinet, not included)