Applications
- Controller/transmitter for 3S probes
Description
The 3S-PC1000 is a controller/transmitter compatible with all 3S Analyzers online probes. The controller is able to manage the simultaneous analysis of two different probes. To know more about 3S probes, please visit the corresponding page.
Method
The method is based on fluorescence spectroscopy of organic species dissolved in water.
Reagents
No reagents required.
Features
Range | 30 ppm (eq. PhOH) |
Wavelength (ex.) | 280 nm |
Wavelength (em.) | 300 - 400 nm |
The principle of measurement is based on the determination of the absorbance of organic species dissolved in water. The low range version of the probe is indicated for organic loads with a COD(KHPeq) between 0 and 370 mg/l.
Reagents
No reagents required.
Features
Wavelength | 254 nm |
The principle of measurement is based on the determination of the absorbance of organic species dissolved in water. The high range version of the probe is indicated for organic loads with a COD(KHPeq) between 0 and 1000 mg/l.
Reagents
No reagents required.
Features
Wavelength | 254 nm |
Technical Specifications
Model | 3S-PC1000 |
Sampling Mode | Continuous. |
Sample | Pressure: Atmospheric Temperature: 5 ‐ 45°C (41 ‐ 113°F) Flow Rate: 80 to 500 mL/min Connection: 6 mm (¼-in.) |
Drain |
Pressure-free drain Connection: 12 mm (½-in.) |
Input Streams | 2 |
Material | Epoxy-coated stainless steel AISI 304 |
Dimensions (H x W x D) | 300 x 380 x 210 mm (11.7 x 14.8 x 8.2 in) |
Weight | 10 kg (22 lbs) |
Power Supply |
Input Voltage: 115 VAC,230 VAC Power consumption: max. 10 W |
Output signals | Modbus via Ethernet |
Alarms | 4 SPDT programmable voltage-free relays |
Digital Input | Online |
Ambient Temperature | 5 - 50 °C (41 - 122 °F) |
Ambient Humidity | 10 - 90 % RH (non-condensing) |
Protection Grade | IP65 (indoor only, outdoor use possible with external cabinet, not included) |